Table 2.

Performance of diagnostic tests approved by ANVISA in Brazil for COVID-19. (*data extracted from technical instructions from the diagnostic kits available in the register area ate ANVISA website; NA: not available; CI confidence interval).

Test Name Accuracy data* Methodology Sample Result time
One Step COVID-2019 Test Sensibility: 86,43% (CI 95%: 82,41-89,58)
Specificity: 99,57% (CI 95%: 97,63 - 99,92%)
Immunochromatographic qualitative assay for detection of both IgG and IgM anti-coronavirus Whole blood, serum or plasma (10 µL) 15min
CORONAVÍRUS RAPID TEST Sensibility: 86,4% (CI 95%: 82,41% - 89,58%)
Specificity: 99,57% (CI 95%: 97,63% - 99,92%)
Immunochromatographic qualitative assay for detection of both IgG and IgM anti-coronavirus Whole blood, serum or plasma (10 µL) 15min
CORONAVÍRUS IgG/IgM (COVID-19) IgG
Sensibility: >99,9%
Specificity: 98,0%
Accuracy: 98,6%
IgM
Sensibility: 85,0%
Specificity: 96,0%
Accuracy: 92,9%
Immunochromatographic qualitative assay for detection of both IgG and IgM anti-coronavirus Whole blood, serum or plasma (10 µL) 10-20 min
MedTeste Coronavírus (COVID-19) IgG/IgM (TESTE RÁPIDO) IgG
Sensibility: 97,4% (CI 95%: 86,2% - 99,9%)
Specificity: 99,3% (CI 95%: 96,2% - 99,9%)
Accuracy: 98,9% (CI 95%: 96,1% - 99,9%)
IgM
Sensibility: 86,8% (CI 95%: 71,9% - 95,6%)
Specificity: 98,6% (CI 95%: 95,0% - 99,8%)
Accuracy: 96,1% (CI 95%: 92,2% - 98,4%)
mmunochromatographic qualitative assay for detection of both IgG and IgM Whole blood, serum or plasma (10 µL) NA
Família Teste Rápido em Cassete 2019-nCoV IgG/IgM (sangue total/soro/plasma) IgG
Sensibility: 100% (CI 95%: 86.0% - 100%)
Specificity: 98.0% (CI 95%: 89.4% - 99.9%)
Accuracy: 98.6% (CI 95%: 92.3% - 99.96%)
IgM
Sensibility: 85.0% (CI 95%: 62.1% - 96.8%)
Specificity: 96.0% (CI 95%: 86.3% - 99.5%)
Accuracy: 92.9% (CI 95%: 84.1% - 97.6%)
Immunochromatographic qualitative assay for detection of both IgG and IgM against 2019-nCoV Whole blood, serum or plasma (10µL) 10 min