Table 4.

Zero-frequency DQE for a 120-kVp spectrum and 300-mm water attenuation, for Si and CdTe detector designs, with scatter from the imaged object and antiscatter grid. The 1-D grid of 0.1-mm thick lamellae with 1-mm spacing was used for all detector designs.

Detector design Water detection Iodine detection Water quantification Iodine quantification
60-mm Si, W 0.52 0.39 0.29 0.23
60-mm Si, no W 0.55 0.45 0.40 0.33
3 mm CdTe, 500  μm pixels 0.73 0.49 0.33 0.22
1.6-mm CdTe, 225  μm pixels, σ=22.7  μm 0.65 0.40 0.16 0.10
1.6-mm CdTe, 225  μm, σ=12.1  μm 0.67 0.43 0.20 0.13