TABLE 3.

Linear mixed-effects model analysis of PDFF bias for five different 2D Cartesian sequential CSE-MRI acquisitions.

LFA-linear HFA-linear LFA-centric HFA-centric VFA-centric
Intercept 0.02±0.71
p = 1.00
−0.75±0.88
p = 0.39
0.15±0.71
p = 0.83
0.12±0.68
p = 0.86
0.21±0.71
p = 0.77
true PDFF −0.05±0.03
p = 0.10
−0.17±0.04
** p < 0.01
−0.04 ±0.03
p = 0.14
−0.09 ±0.03
** p < 0.01
−0.06±0.03
p = 0.05
T1,water
(×10−3)
0.80±0.83
p = 0.34
2.47±1.04
* p = 0.02
0.66±0.83
p = 0.43
1.10±0.80
p = 0.17
0.56±0.83
p = 0.50
true PDFF· T1,water
(×10−3)
0.06±0.03
p = 0.09
0.30±0.04
** p < 0.01
0.04 ±0.03
p = 0.23
0.07±0.03
* p = 0.04
0.06±0.03
p = 0.09
*

p<0.05

**