Skip to main content
. 2020 Jul 17;130:104549. doi: 10.1016/j.jcv.2020.104549

Fig. 1.

Fig. 1

Test results of examined SARS-CoV-2 serology assays.

A-E The scatter plots with logarithmic scales display test results of SARS-CoV-2 PCR positive patients (blue dots, n = 75) and pre-pandemic blood donors (green dots, n = 320). Each dot represents one sample. Cut-off values for positivity are 1.1 for Euroimmun, 15 for Diasorin and 1 for Roche, Wantai and Siemens (upper dotted line). Cut-off values for borderline test results are 0.8 for Euroimmun and 12 for Diasorin (lower dotted line). F Test results of SARS-CoV-2 PCR positive patients (n = 75). Each column represents one patient. Samples are sorted by time since onset of symptoms in days. Negative test results are displayed in grey, positive test results in blue and borderline test results in light blue. Immunosuppressed patients are marked by a * in the bottom row and a ↑ in the bottom row indicates that RT-PCR was performed by an external certified lab.

Abbreviations: A/C.O., absorbance/cut-off value; AU/mL, arbitrary units/mL; COI, cut-off index (signal sample/cut-off); d, days; Dia, Diasorin; Eur, Euroimmun; Roc, Roche; Ratio, ratio (extinction sample/extinction calibrator); Sie, Siemens; Wan, Wantai.