Skip to main content
. 2020 Jul 17;39(12):2373–2377. doi: 10.1007/s10096-020-03983-9

Table 3.

Results of variplex test and RT-PCR assay performed simultaneously in 109 patients (95% CI)

Sensitivity 17.0% (7.6%–30.8%)
Specificity 88.7% (78.1–95.3%)
PPV 53.3% (26.6–78.7%)
NPV 58.5% (47.9–68.6%)
Accuracy 57.8% (48.0–67.2%)