Skip to main content
. 2019 Jul 20;34(4):629–641. doi: 10.1007/s10877-019-00362-4

Table 1.

Individual nociception monitor characteristics

Monitor Manufacturer Autonomic input variables Index Optimal range/cut off
SPI GE Healthcare

Pulse beat interval

Pulse wave amplitude

0–100 20–50
ANI MDoloris Medical Systems Heart rate variability 0–100 50
NOL Medasense

Heart rate and heart rate variability

Pulse wave amplitude

Skin conductance

0–100 10–25
CARDEAN Alpha-2

Blood pressure

Heart rate

0–100 60
Algiscan ID Med Pupil diameter NA Pupil diameter increase > 30%

SPI surgical plethysmographic index, ANI analgesia nociception index, NOL nociception level index, NA not available