Table 3.

SAFRAD reporting criteria [32] 1 and 5 should only be used if 2 and 3 are not available!

Trigger values
Fluoroscopy time  > 60 min

Air kerma area product

(or dose area product)

For cardiac and neurological procedures: > 300 Gy.cm2 or 30,000 cGy.cm2 or 30,000 µGy.m2 or 300,000 mGy.cm2

For other procedures: > 500 Gy.cm2 or 50,000 cGy.cm2 or 50,000 µGy.m2 or 500,000 mGy.cm2

Cumulative air kerma (dose) at the interventional reference point  > 5 Gy or 5000 mGy
Measure peak skin dose  > 3 Gy or > 3000 mGy
Number of cine acquisition series  > 20
Trigger events
Observed radiation injury
Patient had multiple procedures within the last 12 months
Wrong procedure performed
Procedure performed on the wrong patient
Unknown pregnancy at time of procedure