Accuracy, precision, limit of detection (LOD), and limit of quantification (LOQ) of the developed ic-ELISA.

Parameters ic-ELISA GC-FPD
LOD, µg/kg 0.26 (IC15) 1.00
LOQ, µg/kg 0.40–41.8 (IC20–IC80) 2.00
Precision (%CV, n = 4)
2.00 µg/kg 4.60 2.00
1.00 µg/kg 6.70 2.58
0.50 µg/kg 5.50 3.40
Accuracy (%Recovery: n = 5)
2.00 µg/kg 95.3 114.32
1.00 µg/kg 117.8 99.42
0.50 µg/kg 95.72 97.60