Abstract
Two patients with bicuspid aortic valve were selected for aortic valve repair using the Ozaki procedure. Patient-specific models of their aortic roots were generated based on computed tomography data and were 3-dimensional printed using a flexible resin. The models allowed sizing of the valve leaflets and practicing of leaflet suturing. (Level of Difficulty: Advanced.)
Key Words: 3D printing, aortic coarctation, bicuspid aortic valve, computed tomography, Ozaki repair, rapid prototyping, surgical planning, valve repair
Abbreviations and Acronyms: 3D, 3 dimensional; AR, aortic regurgitation; BAV, bicuspid aortic valve; CT, computed tomography; Vmax, maximum velocity
Graphical abstract

Two patients with bicuspid aortic valve were selected for aortic valve repair using the Ozaki procedure. Patient-specific models of their aortic…
Two patients with a diagnosis of bicuspid aortic valve (BAV) were selected for aortic valve repair using the Ozaki procedure. Patient #1 was a 65-year-old woman with functional BAV, asymptomatic moderate-to-severe aortic stenosis, and a small anterior communicating artery aneurysm who was followed up at the University Hospitals Bristol congenital heart disease clinic. Detriment of her functional capacity was observed on cardiopulmonary stress testing. Echocardiographic examination showed severe aortic stenosis, with an aortic valve area of 0.5 cm2, maximum velocity (Vmax) of 4.5 m/s, mean gradient of 45 mm Hg, and normal systolic left ventricular function. The decision was taken to intervene surgically.
Learning Objectives
-
•
To understand the role of 3D printing models in surgical planning for aortic valve repair.
-
•
To appreciate decision-making aspects around aortic valve repair with the Ozaki technique in cases of BAV with different annulus sizes.
Patient #2 was a 34-year-old man with congenital BAV and coarctation of the aorta repaired at the age of 4 years with end-to-end anastomosis and residual hypertension. He presented with symptoms of breathlessness and progression of aortic regurgitation (AR) due to leaflet prolapse. Echocardiographic examination confirmed severe AR, with normal ejection fraction and a dilated left ventricle. Surgical treatment was planned. Detailed patient characteristics are summarized in Table 1.
Table 1.
Patient Characteristics
| Patient #1 | Patient #2 | |
|---|---|---|
| Demographic data | ||
| Age at operation, yrs | 65 | 34 |
| Sex | Female | Male |
| Anatomy | Functional BAV | BAV (LCC-RCC), coarctation of the aorta repair at age 4 yrs |
| Treatment | — | — |
| Surgical indication | Detriment of functional capacity | Progression of AR |
| ECG | SR, 75 beats/min | SR, 56 beats/min |
| Weight, kg | 68 | 91 |
| Height, cm | 176 | 184 |
| BSA, cm2 | 2.2 | 2.6 |
| Hypertension | No | Yes |
| Echocardiogram data | ||
| Aorta Vmax, m/s | 4.5 | 1.9 |
| Peak gradient, mm Hg | 75 | — |
| Mean gradient | 45 | — |
| Aortic regurgitation | Mild | Severe |
| EF, % | 55 | 65 |
| lS′-wave, cm/s | 10 | 11 |
| sS′-wave, cm/s | 8 | 9 |
| E/A | 1.0 | 1.6 |
| E/E′ | 7 | 7.9 |
| LVEDD, mm | 42 | 68 |
| LVESD, mm | 26 | 55 |
| CT data | ||
| Aortic annulus, mm | 24 | 33 × 34 |
| SV, mm | 31 | 40 × 34 |
| Ascending aorta, mm | 38 × 38 | 30 × 32 |
| Descending aorta, mm | 20 | 25 |
| CT acquisition | ||
| Columns, n | 512 | 512 |
| Slice thickness, mm | 0.60 | 0.50 |
| Pixel spacing, mm | 0.39 | 0.34 |
| Cycle time acquisition | End systole | Diastole |
AR = aortic regurgitation; BAV = bicuspid aortic valve; BSA = body surface area; CT = computed tomography; ECG = electrocardiography; EF = ejection fraction; LCC = left coronary cusp; LVEDD = left end-diastolic diameter; LVESD = left end-systolic diameter; RCC = right coronary cusp; SR = sinus rhythm; SV = sinus of Valsalva; Vmax = maximum velocity.
At the multidisciplinary meeting, the decision to carry out aortic leaflet reconstruction using glutaraldehyde-treated autologous pericardium—the Ozaki procedure (1,2)—for both patients was based on the advantages of avoiding anticoagulation (necessary for a mechanical valve prosthesis) and the potentially better longevity of autologous material compared with a biological valve prosthesis. A Ross procedure was also discussed but discounted. Both patients fully accepted the concept of autologous pericardial reconstruction of the aortic leaflets and the available evidence in the literature (2).
As a potential aid to leaflet sizing, patient-specific models of the aortic root for each case were generated based on the patients’ computed tomography (CT) data (Figures 1A and 1B, Videos 1 and 2). Their CT data sets were imported into and processed with commercial software (Mimics, Materialise, Leuven, Belgium) for 3-dimensional (3D) reconstruction (3). A 3D volume of the aortic root was generated (Figures 1C and 1D) and exported to a 3D printer. Aortic root models were printed in house (Form2, Formlabs, Somerville, Massachusetts) by using a soft and resilient compliant compound (Elastic Resin, Formlabs; mechanical properties as per manufacturer’s data sheet: elongation at break: 160%; tensile strength: 3.2 MPa; tear strength: 19.1 kN/m), with a wall thickness of 1 mm. Once manufactured, the models were provided to the surgeon for leaflet sizing and suturing.
Figure 1.
Patient-Specific Models of the Aortic Roots Were Derived From Clinically Indicated Computed Tomography Datasets
(A, B) Segmentation was carried out by using Materialise (Leuven, Belgium) Mimics software to select the area of interest (highlighted in red)—in this case, the aortic root. (C, D) Materialize 3-matic was used to reconstruct the aortic roots in 3 dimensions (3D) and produce an stereolithography file suitable for importing into the 3D printer software. (E, F) The aortic roots were then printed in a flexible resin. Images for Patients #1 and #2 displayed are in the top and bottom rows, respectively.
Online Video 1.
Aortic root model reconstructed from CT data.
Online Video 2.
Aortic root model reconstructed from CT data.
Models were successfully produced for both types of aortic valve disease (Figures 1E and 1F), demonstrating the feasibility of the workflow for pre-sizing aortic valve leaflets in 2 patients with BAV with different underlying causes of valve dysfunction. Models took approximately 1 hour to reconstruct and 6 h to print each. Once provided to the surgeon, the sizing process was successfully performed in the lab (Figure 2A, Table 1, Video 3), and leaflets were cut from GoreTex to practice suturing. Leaflet suturing was also demonstrated to be feasible (Figure 2B), and feedback from the surgeon was extremely positive, highlighting the qualities of the material and the advantage, for prospective cases, of pre-sizing the aortic valve leaflets.
Figure 2.

Examples of Using the Aortic Root Model
Examples of using the model by (A) sizing using the Ozaki sizers and (B) practicing leaflet suturing.
Online Video 3.
Sizing of the aortic sinuses on 3D printed model.
With regard to surgical results, Patient #1 did not present any complications after surgery and was discharged after 5 days. The echocardiogram at 4 weeks post-surgery showed no AR, aortic Vmax of 2.2 m/s, a substantially reduced mean gradient of 10 mm Hg, and aortic valve area of 1.5 cm2. Patient 2 was also discharged 5 days after surgery without complications. An echocardiogram at 4 weeks post-surgery showed no AR, an aortic Vmax of 1.8 m/s, and a reduction in LV dimensions (Table 2).
Table 2.
Post-Operative Results
| Patient #1 | Patient #2 | |
|---|---|---|
| Leaflet sizing | ||
| Leaflet sizes, intraoperative, mm | ||
| RCC | 27 | 27 |
| LCC | 25 | 27 |
| NCC | 27 | 31 |
| Leaflet sizes, model, mm | ||
| RCC | 29 | 33 |
| LCC | 27 | 31 |
| NCC | 33 | 35 |
| Echocardiogram data | ||
| E/A | 1.5 | 1.6 |
| E/E′ | 7.5 | 7.9 |
| LVEDD, mm | 38 | 44 |
| LVESD, mm | 27 | 31 |
NCC = noncoronary cusp; other abbreviations as in Table 1.
3D printing technology is increasingly recognized as a valuable tool for surgical planning, providing an insight into complex intracardiac structures with accurate sizing and providing the surgeon with the ability to visualize the heart before the operation (3). These cases highlight that the Ozaki procedure is a novel scenario in which the possibility of replicating patient-specific anatomies by means of 3D printing technology can be of benefit. The Ozaki procedure is a cutting-edge approach to aortic valve reconstruction with very favorable initial results (4). It is a technique that uses stentless aortic valve replacement and uses autologous pericardium for the reconstruction of the valve leaflets (5). Sizing of the leaflets currently takes place in the surgical theater but, as demonstrated here, could instead be planned ahead, thereby saving time in the operating theater. We note that the pre-sized leaflets were slightly bigger than those ultimately implanted in the patients during surgery, allowing for the surgeon to trim the leaflets once implanted if deemed appropriate/necessary. Furthermore, in reoperation scenarios, when autologous pericardium cannot be used, a pre-operative 3D printing–based aortic valve leaflet using bovine or tissue engineered material could, potentially, help the planning and the execution of the procedure in a patient-specific manner.
The precision of the 3D printing manufacturing process (6,7) enables accurate anatomic replicas to be produced to facilitate patient counseling, offer training opportunities, and inform the clinical decision-making process. In the case of the Ozaki procedure, when suitable imaging data (CT or cardiovascular magnetic resonance) are available, patient-specific aortic root models can be printed in a short time frame, allowing the surgeon to size and prepare the aortic valve leaflets before the actual surgery.
These 2 cases of aortic valve disease with different annulus sizes were successfully reproduced using 3D models, and the short-term outcomes following the Ozaki repair in these patients were excellent, based on an improvement in functional capacity and echocardiography.
Take-Home Message
In light of the experience presented here, further research into 3D printing patient-specific aortic models for surgical planning in valve repair/replacement is certainly warranted, including exploring novel materials (e.g., silicone) compatible with the technology and testing these in a systematic manner. We will also look to expand this study to a larger case series of patients, potentially including longer follow-up.
Footnotes
This work was supported by the Grand Appeal–Bristol Children’s Hospital Charity and grant number CH/17/1/32804 from the British Heart Foundation. The funding sources had no role in the design and conduct of the study; collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and decision to submit the manuscript for publication. The authors have reported that they have no relationships relevant to the contents of this paper to disclose.
The authors attest they are in compliance with human studies committees and animal welfare regulations of the authors’ institutions and Food and Drug Administration guidelines, including patient consent where appropriate. For more information, visit the JACC: Case Reportsauthor instructions page.
Appendix
For supplemental videos, please see the online version of this paper.
References
- 1.Ozaki S., Kawase I., Yamashita H. Reconstruction of bicuspid aortic valve with autologous pericardium. Circ J. 2014;78:1144–1151. doi: 10.1253/circj.cj-13-1335. [DOI] [PubMed] [Google Scholar]
- 2.Ozaki S., Kawase I., Yamashita H., Uchida S., Takatoh M., Kiyohara N. Midterm outcomes after aortic valve neocuspidization with glutaraldehyde-treated autologous pericardium. J Thorac Cardiovasc Surg. 2018;155:2379–2387. doi: 10.1016/j.jtcvs.2018.01.087. [DOI] [PubMed] [Google Scholar]
- 3.Milano E.G., Capelli C., Wray J. Current and future applications of 3D printing in congenital cardiology and cardiac surgery. Br J Radiol. 2019;92:20180389. doi: 10.1259/bjr.20180389. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Vukicevic M., Mosadegh B., Min J.K., Little S.H. Cardiac 3D printing and its future directions. J Am Coll Cardiol Intv. 2017;10:171–184. doi: 10.1016/j.jcmg.2016.12.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Ozaki S., Kawase I., Yamashita H. Aortic valve reconstruction using autologous pericardium for aortic stenosis. Circ J. 2015;79:1504–1510. doi: 10.1253/circj.CJ-14-1092. [DOI] [PubMed] [Google Scholar]
- 6.Witowski J., Wake N., Grochowska A. Investigating accuracy of 3D printed liver models with computed tomography. Quant Imaging Med Surg. 2019;9:43–52. doi: 10.21037/qims.2018.09.16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.George E., Liacouras P., Rybicki F.J., Mitsouras D. Measuring and establishing the accuracy and reproducibility of 3D printed medical models. Radiographics. 2017;37:1424–1450. doi: 10.1148/rg.2017160165. [DOI] [PMC free article] [PubMed] [Google Scholar]