Table 2. Analytical Results of 8-Oxo-dG Detection in the Spiked Urine Samples from our Portable Low-Cost Device Based on the 38-nt and 66-nt Aptasensors.
| spiked 8-oxo-dG (nM) | amount found ± S.D. (nM) | recovery ± RSD (%) |
|---|---|---|
| 38-nt aptasensor | ||
| 0 | 24.2 ± 3.0 | |
| 20 | 43.5 ± 4.9 | 96.3 ± 11.3 |
| 40 | 68.6 ± 2.7 | 110.9 ± 4.0 |
| 66-nt aptasensor | ||
| 0 | 24.3 ± 2.2 | |
| 20 | 48.2 ± 4.0 | 119.7 ± 8.3 |
| 40 | 63.1 ± 3.4 | 97.2 ± 5.4 |