Skip to main content
. 2018 Dec 28;46(2):665–678. doi: 10.1002/mp.13322

Table 3.

Parameters/methods/settings leading to best repeatability for different activity uptake groups

Heterogeneous inserts High uptake spheres Low uptake spheres
Reconstruction Minor impact Minor impact PSF+TOF
Matrix/voxel size Minor impact Minor impact Smaller voxel size (2 × 2 × 2 mm)
Scan duration 300 s 300 s 300 s
Smoothing (FWHM) 6/8 mm 6/8 mm 0 mm
Discretization method FBW (bin width 0.25) FBW (bin width 0.25) FBW (bin width 0.05)
Segmentation method Minor impact Minor impact CT‐based segmentation