Skip to main content
. 2020 Jul 24;15(7):e0236564. doi: 10.1371/journal.pone.0236564

Table 2. Performance of optimized direct RT-qPCR approach with reference to standard approach for detection of SARS-CoV-2 RNA.

Statistic Value 95% CI
Total number of specimens 132 -
True positive 18 -
True negative 112 -
False positive 1 -
False negative 1 -
Sensitivity 95.0% 74% to 99.8%
Specificity 99.0% 95.2% to 99.9%
Accuracy 98.5% 94.6% to 99.8%