Skip to main content
. 2020 Jul 4;32(3):268–273. doi: 10.4103/JOCO.JOCO_76_20

Table 1.

Comparison between the Plusoptix A09, objective retinoscopy, and cycloplegic retinoscopy for spherical, cylindrical, and spherical equivalent values

Total (n=151) Sphere (D) Cylinder (D) Spherical equivalent (D)



Mean±SD 95% CI P Mean±SD 95% CI P Mean±SD 95% CI P
Cycloplegic retinoscopy 0.8±0.82 0.67-0.94 NA 0.34±0.93 −0.55-0.25 NA 0.61±0.81 0.48-0.74 NA
Objective retinoscopy 0.65±0.69 0.54-0.76 NA 0.32±0.86 −0.46-0.18 NA 0.45±0.7 0.38-0.60 NA
Plusoptix 1.12±1.16 0.94-1.31 0.0001 0.83±1.27 −1.04-0.63 0.0001 0.71±1.06 0.53-0.88 0.097

D: Diopter, SD: Standard deviation, CI: Confidence interval, NA: Not available