Table 3.
Polarity | FI | FD | LIFDI | EI | APCI | FAB | DART | ESI | MALDI |
---|---|---|---|---|---|---|---|---|---|
Positive ions | + | + | + | + | + | + | + | + | + |
Negative ions | – | (+)1 | – | – | + | + | + | + | + |
Analyte property | |||||||||
Nonpolar M | + | + | + | + | (+) | + | (+) | – | (+) |
Polar M | + | + | + | + | + | + | + | (+) | + |
Highly polar M | (+) | + | + | (+) | + | + | + | + | + |
Ionic CA | – | + | + | – | (+) | + | + | + | + |
Volatile | + | – | – | + | + | – | + | – | – |
Involatile | – | + | + | – | + | + | + | + | + |
Air/water sensitive | + | – | + | (+)2 | – | (+)2 | – | (+)3 | – |
Other | |||||||||
GC-MS | + | – | (+)4 | + | (+)5 | – | – | – | – |
1Only selected instruments can do; rarely used.
2Special sample handling like glove box required.
3Restricted to ionic analytes in dry aprotic solvents.
4When LIFDI probe is used in FI mode.
5Special GC interface required.