Table 3.

FI, FD, and LIFDI in comparison to other ionization methods.

Polarity FI FD LIFDI EI APCI FAB DART ESI MALDI
Positive ions + + + + + + + + +
Negative ions (+)1 + + + + +
Analyte property
 Nonpolar M + + + + (+) + (+) (+)
 Polar M + + + + + + + (+) +
 Highly polar M (+) + + (+) + + + + +
 Ionic CA + + (+) + + + +
 Volatile + + + +
 Involatile + + + + + + +
 Air/water sensitive + + (+)2 (+)2 (+)3
Other
 GC-MS + (+)4 + (+)5

1Only selected instruments can do; rarely used.

2Special sample handling like glove box required.

3Restricted to ionic analytes in dry aprotic solvents.

4When LIFDI probe is used in FI mode.

5Special GC interface required.