TABLE 2.
CDC-based LDT versus Hologic Panther Fusion EUAa
Sample ID |
CT found using the following test and primer or gene: |
||
---|---|---|---|
UW IDT |
Panther fusion SARS-CoV-2 (EUA), Orf1ab/2ab |
||
N1 | N2 | ||
100 | 30 | 30.4 | 29 |
67 | 20.3 | 20.8 | 19.3 |
76 | 27.1 | 27.6 | 31.1 |
17 | 23.4 | 23.6 | 20.5 |
02 | 18.1 | 17.2 | 19.3 |
59 | 29.4 | 29.8 | 31.7 |
97 | 18.4 | 17 | 18.9 |
52 | 16.8 | 15.9 | 16.9 |
90 | 36.1 | 38.4 | NDET/38.1 |
79 | 38.4 | NDET | NDET/NDET |
All 10 CDC LDT-negative specimens were negative by Hologic Panther Fusion EUA. NDET, not detected.