TABLE 3.
Details of discordant samplesa
Sample ID | SARS-CoV-2 sample-to-answer molecular assay results |
|||
---|---|---|---|---|
Reference method (CT value) |
Xpert Xpress (E/N2 CT values) |
ID NOW | ePlex | |
A-10 | POS (33.1) | POS (32.8/35.8) | POS | NEG |
A-12 | POS (33.2) | POS (31.7/34.6) | NEG | NEG |
A-14 | POS (34) | POS (33.3/35.5) | NEG | NEG |
A-15 | POS (32.6) | POS (32.2/35.4) | NEG | POS |
A-16 | POS (33.2) | POS (33.6/36.4) | NEG | POS |
A-24 | POS (38.5) | NEG (N/A) | NEG | NEG |
A-26 | POS (36.2) | POS (36.6/39.5) | NEG | NEG |
A-103 | POS (32) | POS (31.1/34.3) | NEG | POS |
Discordant sample results are highlighted in bold. CT, cycle threshold; ID, identifier; NEG, negative; POS, positive.