Skip to main content
. 2020 Jul 23;58(8):e00926-20. doi: 10.1128/JCM.00926-20

TABLE 1.

Limits of detection of the Xpert Xpress SARS-CoV-2 test using SARS-CoV-2 in nasopharyngeal-swab clinical matrix

Concn (PFU/ml) No. positive/no. of replicates Mean CT
Hit rate (%)
E N2
0.0200 22/22 35.2 38.9 100
0.0100 22/22 37.1 40.1 100
0.0050 21/22 38.2 41.0 95.5
0.0020 10/22 41.0 41.7 45.4
0.0010 7/22 39.1 41.8 31.8
0.0005 2/22 39.5 41.3 9.1
0.0001 1/22 43.5 42.5 4.5