TABLE 3.
Workflow analysis comparing laboratory-modified CDC COV assay, ACOV assay, and IDNOW assay
Parametera | CDC COV | ACOV | IDNOW |
---|---|---|---|
Off-board lysis | Yes | No | No |
Specimen processing and set up | 1.75 | 1.00 | 0.03 |
Instrument extraction time | 1.30 | 4.0 | 0.05 |
Amplification and real-time detection | 1.25 | 2.25 | 0.22 |
Manual interpretation and result entry | 0.5 | 0.75 | 0.02 |
Total time (h) to resultb | 4.8 | 8.0 | 0.27 |
No. samples processed in 8-h shift per instrument | 58c | 94d | 32e |
Times (hours) per batch for the in-house laboratory-developed test (LDT) and Abbott m2000 assays, and per sample for the ID Now assay.
Times (hours) from sample processing through result reporting.
A maximum of 58 patient samples, not including external positive control, negative control, negative template control, RNase P control.
A maximum of 94 patient samples, not including external positive and negative controls.
Assumes continuous processing of 32 patient samples, all with negative results. Results for positive samples may be generated within 5 min, and results for negative samples are generated within 13 min.