TABLE 1.
Comparison of the SHC-LDT for SARS-CoV-2 and the Accula SARS-CoV-2 PCR test
| SHC-LDTa | Accula SARS-CoV-2 PCR test |
Total | |
|---|---|---|---|
| Detected | Not detected | ||
| Detected | 34 | 16 | 50 |
| Not detected | 0 | 50 | 50 |
| Total | 34 | 66 | 100 |
SHC-LDT, Stanford Health Care laboratory-developed test.