Table 1.

Stability of ANDE desiccant swabs of oral epithelial samples from drinking containers and (3 µL‐dried blood samples on ceramic tiles as assessed by the number of (a) CODIS and (b) FlexPlex27 loci called following Rapid DNA analysis.

a
Sample Type Donor Sample # Time points
1 Month 3 Months 6 Months 9 Months 12 Months
No. of CODIS Loci Called No. of CODIS Loci Called No. of CODIS Loci Called No. of CODIS Loci Called No. of CODIS Loci Called
Oral epithelial 1 1 20 20 20 20 16
2 20 20 20 20 20
2 1 20 20 11 20 20
2 20 20 20 20 20
3 1 20 20 20 12 20
2 20 20 20 20 19
Blood 4 1 20 14 20 20 20
2 20 20 20 20 20
5 1 20 20 20 20 20
2 20 20 20 20 20
6 1 10 20 20 * 20
2 20 20 20 20 20
Average CODIS20 Loci 19.17 19.50 19.25 19.27 19.58
b
Sample Type Donor Sample # Time Points
1 Month 3 Months 6 Months 9 Months 12 Months
No. of FlexPlex Loci Called No. of FlexPlex Loci Called No. of FlexPlex Loci Called No. of FlexPlex Loci Called No. of FlexPlex Loci Called
Oral Epithelial 1 1 27 27 27 26 21
2 26 27 27 27 26
2 1 27 27 15 27 26
2 27 27 27 27 27
3 1 24 23 24 12 24
2 24 23 24 23 24
Blood 4 1 27 20 27 27 27
2 27 27 27 27 27
5 1 27 27 27 27 27
2 27 27 27 27 27
6 1 13 27 27 * 27
2 27 27 27 27 27
Average FlexPlex27 Loci 25.25 25.75 25.50 25.18 25.83
*

No profile generated. The I‐Chip did not fully process the sample; this occurs in approximately 2% of lanes.