Table III.

Comparison of expected and measured HU values for scan C for both scanners (right hand columns). The expected values for scanners A to C are from patient studies. Simulated values are from Monte Carlo simulations of clinical dose distributions. The measured phantom values for scanner 1 and scanner 2 are after a subtraction of 60 HU.

Tissue Published data This study
Mean CT number scanner A 53 (120 kVp) Mean CT number scanner B 54 (120 kVp) Mean CT number scanner C 54 (120 kVp) Simulated value (120 kVp 52 , not specified 55 ) Mean scanner 1 (120 kVp) Mean scanner 2 (120 kVp)
Gray matter 29.9 ± 3.8 33.2 ± 0.74 34.18 ± 0.97 40 52 35.8 ± 9.9 32.9 ± 6.7
White matter 22.7 ± 3.8 25.06 ± 0.6 26.11 ± 0.93 34 52 29.3 ± 9.5 27.4 ± 6.1
Cerebrospinal fluid 13 52 22 ± 10.4 19.5 ± 6.9
Cancellous bone   568–828 55 , 999 52 690 715