Table 3.
Ion chamber (IC) % difference and 3%/2 mm gamma passing rates (mean ± standard deviation) for all measurement devices for IMRT and VMAT plans.
| n | IC (%) | Film | n | ArcCHECK | Delta4 | PD | |
|---|---|---|---|---|---|---|---|
| IMRT | 10 | −0.6 ± 1.0 | 98.1 ± 1.9 | 15 | 99.1 ± 0.7 | 99.0 ± 1.1 | 99.8 ± 0.7 |
| VMAT | 24 | −0.9 ± 1.6 | 96.3 ± 3.5 | 29 | 99.0 ± 0.9 | 99.3 ± 1.2 | 99.9 ± 0.4 |
| ALL | 34 | −0.8 ± 1.4 | 96.8 ± 3.3 | 44 | 99.1 ± 0.9 | 99.2 ± 1.2 | 99.9 ± 0.5 |
| Tolerance Limit | 90.3 | 96.2 | 96.8 | 99.5 | |||
| Action Limit | 86.3 | 96.1 | 95.7 | 98.4 |
IC and film results do not include TG‐119 plans, while TG‐119 plans were measured with ArcCHECK, Delta4, and PD.