. 2020 May 5;21(7):39–48. doi: 10.1002/acm2.12881

Table 3.

Ion chamber (IC) % difference and 3%/2 mm gamma passing rates (mean ± standard deviation) for all measurement devices for IMRT and VMAT plans.

  n IC (%) Film n ArcCHECK Delta4 PD
IMRT 10 −0.6 ± 1.0 98.1 ± 1.9 15 99.1 ± 0.7 99.0 ± 1.1 99.8 ± 0.7
VMAT 24 −0.9 ± 1.6 96.3 ± 3.5 29 99.0 ± 0.9 99.3 ± 1.2 99.9 ± 0.4
ALL 34 −0.8 ± 1.4 96.8 ± 3.3 44 99.1 ± 0.9 99.2 ± 1.2 99.9 ± 0.5
Tolerance Limit     90.3   96.2 96.8 99.5
Action Limit     86.3   96.1 95.7 98.4

IC and film results do not include TG‐119 plans, while TG‐119 plans were measured with ArcCHECK, Delta4, and PD.