Skip to main content
. 2020 Apr 30;21(7):70–76. doi: 10.1002/acm2.12885

Table 4.

Responses to survey questions: planar secondary questions for passing criteria for gamma analysis.

  Gamma distance‐to‐agreement
2 mm 3 mm 4 mm 5 mm
Gamma dose percent difference 1% 0.1% (1)* 0.1% (1) 0.0% (0) 0.0% (0)
2% 5.1% (58)* 0.1% (1) 0.0% (0) 0.0% (0)
3% 4.0% (46)* 84.3% (963) 0.1% (1) 0.1% (1)
4% 0.1% (1) 0.1% (1) 0.7% (8) 0.0% (0)
5% 0.2% (2) 1.0% (12) 0.0% (0) 1.9% (22)

The gamma distance‐to‐agreement ranged from 2 to 5 mm while the gamma dose percent difference ranged from 1 to 5%. Percentage based on number of sites that provided answers for the criteria (N = 1143). *Acceptable criteria based on recommendations for TG‐218