Table 5.

Response to survey questions: 3d/volumetric analysis secondary questions for passing criteria for gamma analysis.

  Gamma distance‐to‐agreement
2 mm 3 mm 4 mm 5 mm
Gamma dose percent difference 1% 0.0% (0)* 0.3% (1) 0.0% (0) 0.0% (0)
2% 5.4% (21)* 0.5% (2) 0.0% (0) 0.0% (0)
3% 3.6% (14)* 84.1% (329) 0.0% (0) 0.5% (2)
4% 0.0% (0) 0.0% (0) 0.8% (3) 0.0% (0)
5% 0.0% (0) 1.5% (6) 0.0% (0) 0.5% (2)

The gamma distance‐to‐agreement ranged from 2 to 5 mm while the gamma dose percent difference ranged from 1 to 5%. Percentage based on number of sites that provided answers for the criteria (N = 391). *Acceptable criteria based on recommendations for TG‐218.