Skip to main content
. 2020 Jul 23;7:893–899. doi: 10.1016/j.toxrep.2020.07.008

Table 2.

Temperature programming of the graphite furnace used to analyze the Cd and Pb in the samples.

Temperature (°C)
Ramp Time (s) Hold Time
(s)
Cadmium Lead
110 110 1 30
130 130 15 30
850 700 10 20
1650 1800 0 5
2450 2450 1 3