Andrews 2012.
| Study characteristics | |||
| Patient Sampling | Case‐control study. Cases were primary angle‐closure suspects (PACS), controls were participants with open‐angles who did not meet the PACS criteria. Data from the right eye were included in the analysis. | ||
| Patient characteristics and setting | Sample size: 442 eyes (370 narrow angle and 72 open angle). Age: mean (SD), 59.8 ± 4.9 years (narrow angle 59.7 0 ± 5.2; controls 60.2 ± 3.2). Sex: 345 (78.0%) female. Setting: secondary care. Country: China. Ethnicity: Chinese. Exclusions: prior intraocular surgery, excessively high risk of acute angle‐closure attack, cataract preventing anterior chamber imaging. |
||
| Index tests |
LACD: graded as a percentage fraction of adjacent corneal thickness at the temporal limbus: >100%, 75%, 40%, 25%, 15%, 5%, and 0%, cut‐off value used ≤ 25%. SPAC: measurements ranged from 1 to 12, cut‐off value used ≤ 6. |
||
| Target condition and reference standard(s) | PACS: participants with pigmented trabecular meshwork not visible in at least two quadrants (≥180 degrees) on gonioscopy (without PAS, glaucomatous optic neuropathy or elevated IOP). | ||
| Flow and timing | There were no uninterpretable test results or exclusions reported. The index test and reference standard were conducted on the same occasion. | ||
| Comparative | |||
| Notes | Conflicts of interest: Dr Kashiwagi has a Japanese patent on the SPAC (Japanese patent No. 3878164). Dr Friedman currently has the SPAC instrument on loan from Carl Zeiss Meditec. | ||
| Methodological quality | |||
| Item | Authors' judgement | Risk of bias | Applicability concerns |
| DOMAIN 1: Patient Selection | |||
| Was a consecutive or random sample of patients enrolled? | Yes | ||
| Was a case‐control design avoided? | No | ||
| Did the study avoid inappropriate exclusions? | Yes | ||
| Could the selection of patients have introduced bias? | High risk | ||
| Are there concerns that the included patients and setting do not match the review question? | High | ||
| DOMAIN 2: Index Test (LACD) | |||
| Were the index test results interpreted without knowledge of the results of the reference standard? | Yes | ||
| If a threshold was used, was it pre‐specified? | Yes | ||
| Could the conduct or interpretation of the index test have introduced bias? | Low risk | ||
| Are there concerns that the index test, its conduct, or interpretation differ from the review question? | Low concern | ||
| DOMAIN 2: Index Test (Scheimpflug photography) | |||
| DOMAIN 2: Index Test (AS‐OCT) | |||
| DOMAIN 2: Index Test (SPAC) | |||
| Were the index test results interpreted without knowledge of the results of the reference standard? | Yes | ||
| If a threshold was used, was it pre‐specified? | Yes | ||
| Could the conduct or interpretation of the index test have introduced bias? | Low risk | ||
| Are there concerns that the index test, its conduct, or interpretation differ from the review question? | Low concern | ||
| DOMAIN 2: Index Test (Flashlight) | |||
| DOMAIN 3: Reference Standard | |||
| Is the reference standards likely to correctly classify the target condition? | Yes | ||
| Were the reference standard results interpreted without knowledge of the results of the index tests? | Yes | ||
| Could the reference standard, its conduct, or its interpretation have introduced bias? | Low risk | ||
| Are there concerns that the target condition as defined by the reference standard does not match the question? | Low concern | ||
| DOMAIN 4: Flow and Timing | |||
| Was there an appropriate interval between index test and reference standard? | Yes | ||
| Were all patients included in the analysis? | Yes | ||
| Did all patients receive a reference standard | Yes | ||
| Could the patient flow have introduced bias? | Low risk | ||