Skip to main content
. 2020 May 29;2020(5):CD012947. doi: 10.1002/14651858.CD012947.pub2

Andrews 2012.

Study characteristics
Patient Sampling Case‐control study. Cases were primary angle‐closure suspects (PACS), controls were participants with open‐angles who did not meet the PACS criteria. Data from the right eye were included in the analysis.
Patient characteristics and setting Sample size: 442 eyes (370 narrow angle and 72 open angle).
Age: mean (SD), 59.8 ± 4.9 years (narrow angle 59.7 0 ± 5.2; controls 60.2 ± 3.2).
Sex: 345 (78.0%) female.
Setting: secondary care.
Country: China.
Ethnicity: Chinese.
Exclusions: prior intraocular surgery, excessively high risk of acute angle‐closure attack, cataract preventing anterior chamber imaging.
Index tests LACD: graded as a percentage fraction of adjacent corneal thickness at the temporal limbus: >100%, 75%, 40%, 25%, 15%, 5%, and 0%, cut‐off value used ≤ 25%.
SPAC: measurements ranged from 1 to 12, cut‐off value used ≤ 6.
Target condition and reference standard(s) PACS: participants with pigmented trabecular meshwork not visible in at least two quadrants (≥180 degrees) on gonioscopy (without PAS, glaucomatous optic neuropathy or elevated IOP).
Flow and timing There were no uninterpretable test results or exclusions reported. The index test and reference standard were conducted on the same occasion.
Comparative  
Notes Conflicts of interest: Dr Kashiwagi has a Japanese patent on the SPAC (Japanese patent No. 3878164). Dr Friedman currently has the SPAC instrument on loan from Carl Zeiss Meditec.
Methodological quality
Item Authors' judgement Risk of bias Applicability concerns
DOMAIN 1: Patient Selection
Was a consecutive or random sample of patients enrolled? Yes    
Was a case‐control design avoided? No    
Did the study avoid inappropriate exclusions? Yes    
Could the selection of patients have introduced bias?   High risk  
Are there concerns that the included patients and setting do not match the review question?     High
DOMAIN 2: Index Test (LACD)
Were the index test results interpreted without knowledge of the results of the reference standard? Yes    
If a threshold was used, was it pre‐specified? Yes    
Could the conduct or interpretation of the index test have introduced bias?   Low risk  
Are there concerns that the index test, its conduct, or interpretation differ from the review question?     Low concern
DOMAIN 2: Index Test (Scheimpflug photography)
DOMAIN 2: Index Test (AS‐OCT)
DOMAIN 2: Index Test (SPAC)
Were the index test results interpreted without knowledge of the results of the reference standard? Yes    
If a threshold was used, was it pre‐specified? Yes    
Could the conduct or interpretation of the index test have introduced bias?   Low risk  
Are there concerns that the index test, its conduct, or interpretation differ from the review question?     Low concern
DOMAIN 2: Index Test (Flashlight)
DOMAIN 3: Reference Standard
Is the reference standards likely to correctly classify the target condition? Yes    
Were the reference standard results interpreted without knowledge of the results of the index tests? Yes    
Could the reference standard, its conduct, or its interpretation have introduced bias?   Low risk  
Are there concerns that the target condition as defined by the reference standard does not match the question?     Low concern
DOMAIN 4: Flow and Timing
Was there an appropriate interval between index test and reference standard? Yes    
Were all patients included in the analysis? Yes    
Did all patients receive a reference standard Yes    
Could the patient flow have introduced bias?   Low risk