Skip to main content
. 2019 Dec 18;58(3):253–259. doi: 10.1007/s10840-019-00685-x

Table 2.

Endotak Reliance calcification with and without Gore ePTFE-covered shocking coils

With Gore Without Gore
Number 102 7
Median implant time (IQR), yrs 10.4 (4.3, 11.7) 7.3 (5.5, 8.9)*
Sign of Failure
  High shocking impedance 66 -
  High pacing impedance 31 7
  High shocking and pacing impedance 3 -
Location of calcification
  Proximal shocking coil 4 -
  Distal shocking coil 25 -
  Pacing electrode 33 4
  Both shocking coils 25 -
  One or both coils & pacing electrode 14 3

*p = 0.36

IQR = interquartile ranges