Table 1.
TECHNICAL SPECIFICATIONS |
CONSIP S.p.A RECOMMENDED INTENSIVE CARE VENTILATORS’ SPECIFICATIONS |
NHS RECOMMENDED INTENSIVE CARE VENTILATORS’ SPECIFICATIONS |
WHO RECOMMENDED INTENSIVE CARE VENTILATORS’ SPECIFICATIONS |
---|---|---|---|
ISO 80601-2-12:2011 COMPLIANT | Required | Required | |
PATIENT TYPE | Adult, pediatric | Adult | Adult, pediatric |
CONTROLS/SETTING RANGES | |||
Tidal volume, mL | Required at least one setting of 400 ml, Preferred: 350–450 or 250–600 or up to 800 | 20–2000 | |
Respiratory rate, breaths/min | 10–30 | 10–60 | |
Trigger mechanism | pressure, flow with high sensitivity (>0,3 l/min) | Required | |
FiO2, % | 21–100 | Required at least 50% or 60% and 100% options. Preferred 30–100 (35–80% for CPAP) | 21–100 |
Inspiratory flow rate, L/min | 0–200 | 0–100 | 1–160 |
Inspiratory pressure, cm H2O | 15–40 | 0–40 | |
IE ratio | Adjustable |
Required. 1:2 (i.e. expiration lasts twice as long as inspiration) Preferred: adjustable (1:1–1:3) |
Required |
PEEP/CPAP, cm H2O | 5–20 (5–15 for CPAP) | 0–20 | |
Pressure support, cm H2O | Required | 0–35 (max 70) | 5–20 adjustable |
Leak compensation | Required | ||
INVASIVE and NON INVASIVE VENTILATION MODES | |||
CMV - volume controlled (VCV) | Required | Required | Required |
Volume assist/control mode (VCAC) | Required | ||
Synchronized Intermittent Mandatory Ventilation (SIMV) | Preferred | Required | |
CMV- pressure controlled (PCV) | Required | Required | Required |
Pressure support mode (PSV) | Required | Required | |
Pressure assist/control mode (PCAC) | Required | ||
Pressure Regulated Volume Controlled (PRVC) | Required | Required | |
Non-invasive ventilation (CPAP, BIPAP) | Required | Required | Required |
High-frequency ventilation | Preferred | ||
PATIENT ASSESSMENT TOOLS | |||
Maximum waveforms displayed | At least 3 waveforms displayed at the same time | At least 3 | |
Lung recruitment tools (PV loops) | At least 2 loops at the same time | At least 3 | |
Capnography/CO2 monitoring | Required | Preferred | |
Other patient assessment tools | Required: Endotracheal and tracheostomy tube compensation | ||
INTEGRATED CAPABILITIES | |||
Integrated nebulizer | Required | ||
Other integrated capabilities | Preferred inlet gas supply (O2), | Required. inlet gas supply (O2) pressure range 35 psi to 65 ps. | |
MONITORED/DISPLAYED PARAMETERS | |||
Peak inspiratory pressure | Required | Required | |
Airway pressure | Required | Required | |
PEEP pressure | Required | Required | Required |
Tidal volume | Required | Required | Required |
Minute volume | Required | Required | |
Spontaneous minute volume | Required | ||
FiO2 (analyzed %) | Required | Required | Required |
Respiratory rate | Required | Required | Required |
Resistance | Required | ||
Static and dynamic Compliance | Required | ||
IE ratio | Required | ||
Others | Continuous high pressure/occlusion, Required | Required: Ventilation mode | |
PATIENT ALARMS | |||
Low/high FiO2 | Visual and audible Required | Visual and audible Required | |
Low minute volume | Visual and audible Required | Visual and audible Required | |
High minute volume | Visual and audible Required | Visual and audible Required | |
Low inspiratory pressure | Required | Required | Visual and audible Required |
High pressure | Required | Required | Visual and audible Required |
Loss of PEEP | Required | ||
Apnea | Required | Visual and audible Required (adjustable) | |
Continuous high pressure/occlusion | Visual and audible Required | ||
Inverse IE ratio | |||
High respiratory rate | Visual and audible Required | Visual and audible Required | |
High PEEP | Required | Visual and audible Required | |
Breathing circuit disconnect | Required | required | Visual and audible Required |
High/low Tidal Volume | Required | Required | Required |
EQUIPMENT ALARMS | |||
Gas supply failure | Visual and audible Required | Required | Visual and audible Required |
Power failure | Visual and audible Required | Required | Visual and audible Required |
Vent inoperative | Required | Visual and audible Required | |
Low battery | Required | Visual and audible Required | |
Self-diagnostics | Visual and audible Required | ||
DISPLAY | |||
Type | touch screen | ||
Size, cm (in) | >30 (12) | ||
PATIENT TRANSPORT CAPABILITY | Required | Required | |
Optional equipment required for patient transport | Cart | Cart | |
ON-BOARD AIR COMPRESSOR OR TURBINE | Required | Required Air Compressor, Air turbine is an alternative | |
INTERNAL BACK-UP BATTERY | Required | Required | Required |
Operating time, hr | >30 min | >20 min | ≥1 |