Table 1. Summary of the LC-MS/MS assay validation data provided by kit manufacturer.

Analyte, unit
(polarity: MRM1; MRM2)
Reagent/
Manufacturer
Assay range Limit of
quantification
Inter-assay
precision, CV%
Recovery, %
Aldosterone, nmol/L
(-:359®331;359®189)
MassChrom Steroids in Serum and Plasma/
Chromsystems
0.039–16.62 0.039 5.9% 95%
Cortisol, µmol/L
(+:363®97;363®121)
0.004–1.44 0.004 5.0% 93%
DHEAS, µmol/L
(-:369®99;369®82)
0.066–48.78 0.066 6.9% 90%
Testosterone, nmol/L
(+:289®97;289®109)
0.02–83.28 0.02 8.9% 95%
Progesterone, nmol/L
(+:315®97;315®109)
0.095–79.5 0.095 10.2% 82%
OHP, nmol/L
(+:331®109;331®97)
0.12–90.9 0.12 9.5% 92%
Both MRM channels were validated using two different LC-MS/MS instruments. LC-MS/MS - liquid chromatography coupled to the tandem mass spectrometry. MRM - multiple reaction monitoring. DHEAS - dehydroepiandrosterone sulfate. OHP - 17-hydroxyprogesterone.