Skip to main content
. 2020 Aug 5;30(3):030701. doi: 10.11613/BM.2020.030701

Table 2. Summary of the immunoassay validation data provided by kit manufacturers.

Analyte, unit Reagent/
Manufacturer
Assay range Limit of
quantification
Total precision, CV % Recovery, %
Aldosterone, nmol/L Liaison Aldosterone/
DiaSorin
0.027–2.77 0.053 9.5% 91%
Cortisol, µmol/L Alinity i Cortisol/
Abbott
0.028–1.650 0.028 3.7% n.a.
DHEAS, µmol/L Architect DHEA-S/
Abbott
0.08–40.71 n.a. 7.41% 102%
Testosterone, nmol/L Architect 2nd gen
Testosterone/Abbott
0.13–64.57 0.15 4.6% n.a.
Progesterone, nmol/L Alinity i Progesterone/
Abbott
1.6–127.2 1.6 5.8% n.a.
Architect Progesterone/
Abbott
n.a.–127.2 n.a. 6.2% 96%
OHP, nmol/L 17-OH-progesterone/
DIAsource
0.10–60.52 n.a. n.a. 105%
DHEAS - dehydroepiandrosterone sulfate. OHP - 17-hydroxyprogesterone. n.a. = not available