Figure 8.

Figure 8

Experimental results of impact of glucose concentration variation in samples with DI water + serum + saline solution. (a) Frequency shift of the sensor as the response to alternatively changing the glucose concentration from zero to 200 mM/l. It could be seen that, the proposed sensor presents a stable and repeatable response over time. (b) Frequency shift of the sensor as the response of small variation of glucose concentration. It could be seen that, according to introducing of the skin between the sensor and the sample, the overall sensitivity is reduced to 38 kHz/1 mM/l of glucose concentration variation.