Table 2.
Limit of Detection Verification Panel Results for the Abbott RealTime SARS-CoV-2 Assay Compared to the UW Virology SARS-CoV-2 CDC-based Laboratory Developed Test.
Target Concentration (copies/mL) | Number of Replicates Tested | Number of Replicates Detected Abbott Assay | Number of Replicates Detected or Inconclusive UW Virology LDT |
---|---|---|---|
100 | 20 | 20 | 16 |
50 | 20 | 19 | 8 |
25 | 20 | 16 | 3 |
10 | 40 | 14 | 4 |
5 | 20 | 4 | 0 |