Skip to main content
. 2020 May 28;129:104474. doi: 10.1016/j.jcv.2020.104474

Table 2.

Limit of Detection Verification Panel Results for the Abbott RealTime SARS-CoV-2 Assay Compared to the UW Virology SARS-CoV-2 CDC-based Laboratory Developed Test.

Target Concentration (copies/mL) Number of Replicates Tested Number of Replicates Detected Abbott Assay Number of Replicates Detected or Inconclusive UW Virology LDT
100 20 20 16
50 20 19 8
25 20 16 3
10 40 14 4
5 20 4 0