Table 1.

Limits of detection and quantification of the 1:7 dilution protocol in HIV-1 RNA controls

APT-Quant-dil
Reference Value mean SD Quantified Detected
(log10 cop/mL) (log10 cop/mL) (%) (%)
1.56 0 0 0 80
1.89 0 0 0 93
2.33 2.43 0.1 33 100
2.49 2.50 0.12 60 100
2.56 2.51 0.12 67 100
2.62 2.61 0.11 93 100
2.76 2.75 0.18 100 100
2.93 2.89 0.12 100 100

Quantified results from diluted HIV-1 RNA controls tested in triplicate with APT-Quant-std to determine the reference value and tested in 15 replicates with APT-Quant-dil to determine percent quantified and detected, and the mean and standard deviation (SD).