Table 1.

Protocol parameters for (left) the free‐breathing radial Dixon‐RAVE and (right) the breath‐hold Cartesian reference acquisition

Parameters Field strength: 3 T
Protocol 1: Radial Protocol 2: Cartesian
Sequence Dixon‐RAVE VIBE
Number of echoes 6 6
First echo TE1 (ms) 1.23 1.09
Echo spacing ΔTE (ms) 1.23 1.23
Repetition time TR (ms) 8.75 9
Flip angle
5
4
Field‐of‐view ( mm3) 410 × 410 × 260 332 × 380 × 192
Base resolution (pixels) 256 × 256 160 × 140
Pixel size ( mm2) 1.6 × 1.6 1.2 × 1.2 (interp.)a
Slice thickness (mm) 5 6 (acquired); 3 (interp.)c
Slice resolution (%) 100 50
Readout mode bipolar bipolar
Bandwidth (Hz/px) 1090 1080
Number of slices 48‐64 64‐96
Averages 1 1
Radial views 400
Radial sampling Golden‐angle
Total measurement time 3 minutes 17 second‐4  minutes 13 seconds 14 seconds‐20 secondsb

Note: For the radial acquisition, the total measurement time includes the gradient calibration time.

a

Acquired pixel size: 2.4 × 2.4 mm.

b

4‐fold accelerated using CAIPIRINHA (twofold acceleration along ky as well as kz).

c

The Cartesian protocol applied a “slice resolution" of 50%, ie, only 50% of the phase encoding steps are acquired and zero‐padded, resulting in 2 times higher nominal slice resolution (interpolated).