Table 2. Comparison of two measurements with the AS-OCT and Pentacam of the same eye.
| Variable | Mean difference (mean ± SD) |
COR | Relative COR, % | LLOA | ULOA | LOA range | R2 | P |
|---|---|---|---|---|---|---|---|---|
| Central CT (μm) | 9.64±6.68 | 13.10 | 2.42 | −3.46 | 22.74 | 26.21 | 0.005 | 0.639 |
| Superior CT (μm) | 25.22±26.31 | 51.57 | 7.87 | −26.35 | 76.79 | 103.14 | 0.003 | 0.717 |
| Inferior CT (μm) | 14.66±24.51 | 48.06 | 7.80 | −33.40 | 62.72 | 96.12 | 0.022 | 0.314 |
| Nasal CT (μm) | 29.86±29.85 | 56.21 | 8.85 | −26.35 | 86.07 | 112.43 | 0.014 | 0.425 |
| Temporal CT (μm) | 22.04±24.32 | 47.69 | 7.83 | −25.65 | 69.73 | 95.37 | 0.015 | 0.410 |
| ACD (mm) | −0.075±0.15 | 0.294 | 9.26 | −0.37 | 0.22 | 0.588 | 0.013 | 0.430 |
COR, coefficient of repeatability; LLOA, lower limit of agreement; ULOA, upper limit of agreement; CT, corneal thickness; ACD, anterior chamber depth.