. 2020 Jul;8(14):857. doi: 10.21037/atm-20-187

Table 2. Comparison of two measurements with the AS-OCT and Pentacam of the same eye.

Variable Mean difference
(mean ± SD)
COR Relative COR, % LLOA ULOA LOA range R2 P
Central CT (μm) 9.64±6.68 13.10 2.42 −3.46 22.74 26.21 0.005 0.639
Superior CT (μm) 25.22±26.31 51.57 7.87 −26.35 76.79 103.14 0.003 0.717
Inferior CT (μm) 14.66±24.51 48.06 7.80 −33.40 62.72 96.12 0.022 0.314
Nasal CT (μm) 29.86±29.85 56.21 8.85 −26.35 86.07 112.43 0.014 0.425
Temporal CT (μm) 22.04±24.32 47.69 7.83 −25.65 69.73 95.37 0.015 0.410
ACD (mm) −0.075±0.15 0.294 9.26 −0.37 0.22 0.588 0.013 0.430

COR, coefficient of repeatability; LLOA, lower limit of agreement; ULOA, upper limit of agreement; CT, corneal thickness; ACD, anterior chamber depth.