Table 2.

Issues related to safely performing pulmonary function test: spirometry manufacturers’ proposed solutions.

Issue Proposed solutions
Vitalograph26 Morgan Scientific28 ndd27 Vyaire29
Cleaning and infection control New BVF for each pt
Clean the exterior surface with a 70% isopropyl alcohol solution
The interior of the patient circuit requires no decontamination between tests
If internal contamination is suspected, follow appropriate protocol
New BVF for each pt
Clean the exterior surface with a 70% isopropyl alcohol solution
The interior of the patient circuit requires no decontamination between tests
Equipment cleaning and disinfection required after use on infected subjects or prior to use on ID
The ndd hygiene solution, which uses the inserts spirette, FlowTube and bamette, requires no cleaning of internal tubing or sensor
Surface cleaning is required
Separate test / administrative area
Risk assessment questionnaire
Use MicroGard II BVF
Minimize contamination performing PFTs
Use an enzyme cleaner with neutral pH (pH 6–8)
Do not use temperatures above 130 °F
HCWs protection PPE:
- surgical mask
- disposable gloves
Minimum wait time between patients 5 minutes
Ventilation Air ventilation and sterilization
Critical issue Measurements are influenced by the filter's resistance

List of Abbreviations: BVF: Bacterial Viral Filter; HCWs: health-care workers; ID: Immunocompromised patients; PFTs: Pulmonary Function Tests; PPE: personal protective equipment; Pt: patient.