Table 2.
Issue |
Proposed solutions |
|||
---|---|---|---|---|
Vitalograph26 | Morgan Scientific28 | ndd27 | Vyaire29 | |
Cleaning and infection control | New BVF for each pt Clean the exterior surface with a 70% isopropyl alcohol solution The interior of the patient circuit requires no decontamination between tests If internal contamination is suspected, follow appropriate protocol |
New BVF for each pt Clean the exterior surface with a 70% isopropyl alcohol solution The interior of the patient circuit requires no decontamination between tests Equipment cleaning and disinfection required after use on infected subjects or prior to use on ID |
The ndd hygiene solution, which uses the inserts spirette, FlowTube and bamette, requires no cleaning of internal tubing or sensor Surface cleaning is required Separate test / administrative area Risk assessment questionnaire |
Use MicroGard II BVF Minimize contamination performing PFTs Use an enzyme cleaner with neutral pH (pH 6–8) Do not use temperatures above 130 °F |
HCWs protection | PPE: - surgical mask - disposable gloves |
|||
Minimum wait time between patients | 5 minutes | |||
Ventilation | Air ventilation and sterilization | |||
Critical issue | Measurements are influenced by the filter's resistance |
List of Abbreviations: BVF: Bacterial Viral Filter; HCWs: health-care workers; ID: Immunocompromised patients; PFTs: Pulmonary Function Tests; PPE: personal protective equipment; Pt: patient.