Matrix-matched calibration |
External calibration presupposes the preparation of several samples from blank matrix spiked at different analyte concentrations before injection with linear calibrations calculated for each analyte. |
|
Laborious
Time-consuming in relation to the necessity of processing of the blank matrix as the sample
The test sample composition variance must be small
Limitation in bioavailability of appropriate blanks
|
[46,57] |
Isotope labeled internal standard |
The use of internal standard implies the use of a substance with identical or similar ionization properties and very close retention time to that of the analyte. According to literature 13C-labelled IS mimes better than 2H-labelled IS the target analyte. |
|
Expensive if SIL-ISs are used.
For many compounds SIL-ISs are not commercially available
Less suitable for multi residual analyses
|
[58,59,60,61,62,63,64,65] |