Skip to main content
. 2020 Jul 3;25(13):3047. doi: 10.3390/molecules25133047

Table 2.

Calibration approaches applied when the blank matrix is available.

Method Theory or Mechanism Advantages Disadvantages References
Matrix-matched calibration External calibration presupposes the preparation of several samples from blank matrix spiked at different analyte concentrations before injection with linear calibrations calculated for each analyte.
  • It permits one to effectively overcome ME using a matrix the same as the sample

  • Laborious

  • Time-consuming in relation to the necessity of processing of the blank matrix as the sample

  • The test sample composition variance must be small

  • Limitation in bioavailability of appropriate blanks

[46,57]
Isotope labeled internal standard The use of internal standard implies the use of a substance with identical or similar ionization properties and very close retention time to that of the analyte. According to literature 13C-labelled IS mimes better than 2H-labelled IS the target analyte.
  • Better assay performance because they show identical behavior to the analyte in sample pretreatment

  • Not time-consuming

  • Particularly useful when homogeneous class of substances are analyzed

  • Expensive if SIL-ISs are used.

  • For many compounds SIL-ISs are not commercially available

  • Less suitable for multi residual analyses

[58,59,60,61,62,63,64,65]