Table 1.
CW | Burst | |
---|---|---|
VMAX, MI | 72.7 Vpp | 73.9 Vpp |
VMAX, TIS | 20.6 Vpp | 68.6 Vpp |
VMAX | 20.6 Vpp | 68.6 Vpp |
MI @VMAX | 0.56 | 1.76 |
TIS @VMAX | 5.92 | 5.93 |
Md @VMAX | 0.062 | 0.154 |
VMAX, MI and VMAX, TIS represent the maximum driving voltages applied to the transducer within MI and TIS safety limits, respectively. VMAX is the maximum driving voltage satisfying both MI and TIS safety limit.
CW continuous wave, MI mechanical index, TIS thermal index in soft tissue, Md modulation depth.