Skip to main content
. 2020 Aug 8;484:112832. doi: 10.1016/j.jim.2020.112832

Table 2.

Precision Results. The table lists the average precision (CV%) across operators (average repeatability) and across days and operators (intermediate precision) across isotypes and high, medium, and low samples. Per EMA guidelines, all results are below a coefficient of variation of 20% (25% for low samples), and often the results are below a 10%.

Samples IgG
IgA
IgM
average repeatability average intermediate precision average repeatability average intermediate precision average repeatability average intermediate precision
High 7.30% 15.30% 7.50% 14.20% 6.88% 16.25%
Med 10.00% 15.70% 8.48% 14.29% 9.96% 18.27%
Low 10.83% 20.65% 10.55% 17.30% 9.29% 24.78%