Result of Reference Range Studies for Qualitative EGFR Assay

Matrix Variant Median Mean current, nA SD 2 SDs 3 SDs Reference range (3 SDs), nA
Plasma T790M −6.43771 −5.96 16.18 32.37 48.55 <43
Plasma L858R −25.84727 −22.79 14.07 28.14 42.21 <20
Plasma Exon 19 −34.24165 −28.05 15.90 31.79 47.69 <20
Saliva T790M 10.49715 8.82 11.58 23.16 34.75 <44
Saliva L858R −0.5126045 −1.13 10.56 21.13 31.69 <31
Saliva Exon 19 −3.927202 −4.46 12.08 24.16 36.24 <32