Contingency Tables and Prediction Performances Achieved by Respective Laboratory in the Ring Trial for the Genomic Allergen Rapid Detection Tiered Approach

BRT
Eurofins
SenzaGen
Reference 1A 1B No Cat 1A 1B No Cat 1A 1B No Cat
1A 8 1 0 10 1 0 10 1 0
1B 0 6 1 0 7 1 3 3 2
No Cat 0 0 8 0 0 8 0 1 8
Accuracy (%) 91.7 92.6 75
Sensitivity 1A (%) 88.9 90.9 90.9
Sensitivity 1B (%) 85.7 87.5 37.5
Sensitivity No Cat (%) 100 100 88.9