-
•
If the laparoscopy is indicated, it should be performed by an experienced surgeon.
-
•
Avoid movement or changing of HCWs during laparoscopy.
-
•
Check all instruments and the proper functioning of the suction system.
-
•
Minimise the amount of Trendelenburg.
-
•
Closed technique can be advisable for obtaining pneumoperitoneum. Incisions for ports must be as small as possible to avoid leakage around ports.
-
•
Close the taps of the trocars before insertion and during the operation. Use balloon trocars and create suitable holes for the intro-duction of leak-free trocars.
-
•
Make a minimum number of incisions, minimum size of incisions, and minimum exchange of the instruments.
-
•
If a patient is having COVID-19 or is suspected, the operation is performed in a negative pressure environment; keep pressure difference between the operating room below - 4.7 Pa.
-
•
CO2 insufflation pressure must be kept to a minimum and an ultrafiltration (smoke evacuation system or filtration) should be used, if available. Set the intraabdominal pressure as low as possible (10–11 mmHg).
-
•
Minimize the use of energy devices, lower the electrocautery power settings; avoid using ultrasonic devices and avoid prolonged desiccation.
-
•
Consider using vacuum suction devices, a closed-circuit smoke evacuation device with a HEPA filter or a ULPA filter if possible.
-
•
Make sure that the taps of the trocars are closed all the time unless evacuation is achieved.
-
•
Any specimen to be removed should also be done at this time of the operation with the abdomen desufflated.
-
•
Fully aspirate the pneumoperitoneum before removing the last trocar through a smoke evacuator device or direct suction.
-
•
After all other ports are removed, ports larger than 5 mm can be closed by a J needle rather than using an Endoclose device, which would increase the risk of gas leaking from the abdomen.
-
•
Ventilate the operating room.
-
•
Manage waste appropriately during and after laparoscopy.