Table 1.

Precision, Accuracy, bias and recovery data for 17OHP in bloodspots by LC-MSMS. * Target value assigned by the CDC. ** LCMSMS method mean (n = 79) from laboratories submitting analytical data to the CDC.

Precision Data (n = 24)
Control 1 Control 2 Control 3
Within-Batch (n = 20)
Mean (nmol/L) 14.6 74.9 155.7
CV% 13.7 7.0 6.2
Between-Batch (n = 22)
Mean (nmol/L) 15.0 74.9 155.1
CV% 11.9 6.1 9.3
Accuracy
Level 1 Level 2 Level 3
CDC Target Value (nmol/L) * 15 66 134
Inter-laboratory Mean (nmol/L) ** 17 82 162
Mean (nmol/L) (n = 10) 19 77 164
Recovery
Level 1 Level 2 Level 3
Baseline Level ND ND ND
Enrichment (nmol/L) 15.2 76.0 152
Mean 15.0 74.9 155
Recovery (%) 101% 98.6% 102%