Table 5.
Previous reports of 3D MRF acquisitions as compared to our proposal.
| Ma et al.10 | Ma et al.12 | Liao et al.7 | Chen et al.36 | Cao et al.11 | Proposed | |
|---|---|---|---|---|---|---|
| Trajectory | Stack-of-spirals | Music generated | Stack-of-spirals | Stack-of-spirals | Spiral projections | Spiral projections |
| Resolution | 1.2 × 1.2 × 3 mm3 | 2.3 Isotropic | 1 mm Isotropic | 1 mm Isotropic | 1/0.8 mm Isotropic | 0.96 Isotropic |
| Spatial coverage | 300 × 300 × 144 mm3 | 300 × 300 × 300 mm3 | 260 × 260 × 192 mm3 | 260 × 260 × 180 mm3 | 240 × 240 × 240 mm3 | 192 × 192 × 192 mm3 |
| Acquisition time | 4.6 min | 37 min | 7.5 min | 6.5–7.1 min | 5/6 min | < 5 min |
| Reconstruction time | 48 min | N.A | 20 h | N.A | 1.5-4 h | < 7 min |