3.
DCT PRO CBCT 4种分辨率模式设置下检测不同深度缺损的Az值
Az values from DCT PRO CBCT scanning with different resolution settings for defects in different depths
| Group | Depth/mm | ||
| 0.1 | 0.2 | 0.3 | |
| Normal quality + normal resolution | 0.405 | 0.545 | 0.665 |
| Normal quality + high resolution | 0.539 | 0.82 | 0.846 |
| High quality + normal resolution | 0.646 | 0.761 | 0.846 |
| High quality + high resolution | 0.681 | 0.853 | 0.865 |