Skip to main content
. 2020 Jul 21;11(9):2019–2027. doi: 10.1007/s13300-020-00879-x
Flash glucose monitoring can improve blood glucose in the short term of 14 days without the need to adjust hypoglycemic treatment.
The estimated A1c used by FreeStyle Libre™ is lower than actual glycated hemoglobin (HbA1c).
The gap between estimated A1c and actual measured HbA1c is positively correlated to the actual HbA1c.
Based on our results, we present an improved model for estimating A1c in flash glucose monitoring.