Skip to main content
. 2020 Jul 27;25(15):3395. doi: 10.3390/molecules25153395

Table 1.

ICP-OES operating conditions.

Parameter Value
Plasma gas flow 10 L/min
Auxiliary gas flow 0.2 L/min
Nebulizer gas flow 0.7 L/min
RF power 1350 watts
Plasma view Axial
Read delay 60 sec
Peristaltic pump flow rate 1.5 mL/min
Spray chamber type Cyclonic
Nebulizer MiraMist
Injector Alumina
Sample tubing Solvaflex 2-stop tubing 0.76 mm i.d.
Drain tubing Solvaflex 2-stop tubing 1.14 mm i.d.
Torch Quartz; single slot
Replicates 3