Skip to main content
. 2020 Jul 22;20(15):4079. doi: 10.3390/s20154079

Table 6.

Relative errors δ for STV and LTV variability indices, determined for original signals FHR_CTG from a fetal monitor output, and for FHR_U from the new Doppler ultrasound signal analysis methods, in relation to the FHR_E reference signal from FECG.

δ (%) FHR Determination Method
FHR_CTG FHR_U D1 FHR_U D2
STV
δS_DAW −66.2 ± 7.99 # −6.4 ± 6.25 −5.0 ± 4.75
δS_YEH −50.8 ± 11.07 −3.0 ± 7.69 −2.4 ± 6.78
δS_HAA −80.0 ± 7.27 −13.3 ± 12.32 −8.9 ± 12.02
δS_ZUG −51.5 ± 9.28 −2.1 ± 9.09 −1.9 ± 8.15
δS_HUE −91.3 ± 8.73 −11.9 ± 16.07 −10.1 ± 12.00
δS_DAL −65.5 ± 7.59 −6.9 ± 6.25 −5.4 ± 4.72
LTV
δL_DAW −10.3 ± 9.00 −0.7 ± 6.47 −0.1 ± 5.76
δL_YEH −6.3 ± 6.78 −0.3 ± 2.79 −0.1 ± 2.36
δL_HAA −2.3 ± 11.16 −1.4 ± 6.89 −0.9 ± 5.09
δL_ZUG −6.7 ± 7.26 −0.2 ± 2.50 −0.1 ± 2.19
δL_HUE 9.80 ± 16.95 −1.3 ± 10.22 −1.3 ± 8.04
δL_DAL −5.9 ± 6.93 −0.1 ± 2.87 −0.1 ± 2.14

# mean value ± SD.