. 2020 Jul 27;25(15):3397. doi: 10.3390/molecules25153397

Table 4.

Validation results of artemether and lumefantrine.

Validation Criteria Artemether Lumefantrine
Linearity Slope 0.9986 1.0342
Intercept 0.0028 0.0001
R2 0.992 0.993
Trueness Level Concentration (mg/mL) Relative Bias (%) Concentration (mg/mL) Relative Bias (%)
80% 0.08 5.03 0.0096 5.22
100% 0.10 −0.08 0.0120 1.72
120% 0.12 0.77 0.0144 0.05
Precision Level Repeatability (RSD%)/Intermediate Precision (RSD%) Repeatability (RSD%)/Intermediate Precision (RSD%)
80% 2.30/2.40 1.75/1.87
100% 1.33/2.86 2.36/3.36
120% 1.18/3.10 0.99/0.99
Accuracy Level β-Expectation Tolerance Limits (%) β-Expectation Tolerance Limits (%)
80% [0.62, 9.43] [1.75, 8.69]
100% [−6.30, 6.15] [−4.92, 8.37]
120% [−6.40, 7.93] [−1.77, 1.87]