Skip to main content
. Author manuscript; available in PMC: 2020 Aug 19.
Published in final edited form as: J Process Control. 2019 Jun 23;80:202–210. doi: 10.1016/j.jprocont.2019.05.017

Table 3:

Time-in-range metrics for the 48-h in silico trial rMPC vs EnMPC.

rMPC EnMPC P
Time <50 mg/dL (%) 0.42 ± 0.98 0.05 ± 0.18 0.0042*
Time <60 mg/dL (%) 1.53 ±2.00 0.25 ± 0.59 0.0000*
Time <70 mg/dL (%) 3.08 ± 3.55 0.78 ± 2.04 0.0000*
Time in range (70, 140) mg/dL 74.05 ± 12.36 71.78 ± 11.41 0.3043
Time in range (70, 180) mg/dL 92.41 ± 8.28 92.81 ± 8.65 0.4239
Time >180 mg/dL (%) 4.51 ± 6.76 6.41 ± 8.16 0.4670
Time >250 mg/dL (%) 0.00 ± 0.00 0.00 ± 0.00 -
Time >300 mg/dL (%) 0.00 ± 0.00 0.00 ± 0.00 -
Variability (SD) 25.4984 ± 9.89 24.4473 ± 10.27 0.5059
Variability (CV) 0.2083 ± 0.08 0.1883 ± 0.08 0.1648
*

(P<0.05)